Device product
Omnera 400T Digital Radiographic System
Z-1331-2022
Product summary
- Firm
- Arcoma AB
- Event
- Event 90446
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 71 systems
- Official record key
device-enforcement:Z-1331-2022
Official wording
Reason: There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
Code information: UDI/DI:07350008750050; 2001-2003, 2006-2044, 2046-2056, 2118-2126, 2128-2131, 2134-2160, 2164-2175, 2177-2195, 2197-2208, 2210-2212, 2214-2224, 2226-2236.
Distribution pattern: US Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA, OR, IN, RI, MI, IL, LA, OH, CO, SC, VA, WA, CA, AZ, NV, AR and the country of Canada.
Derived failure modes
-
Unknown
There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).