Device product

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

Z-1972-2023

May 12, 2023

Class II

Product summary

Firm
Boston Scientific Neuromodulation Corporation
Event
Event 92181
Status
Ongoing
Classification
Class II
Quantity
21,533
Official record key
device-enforcement:Z-1972-2023

Official wording

Reason: Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Code information: Driver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Superion IDS Kit IFU (92479820-02), Surgical Technique Manual (92479821-02)

Distribution pattern: US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    metal fragments