Device product
EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380
Z-1645-2023
Product summary
- Event
- Event 92063
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1421 units US; 3 units OUS
- Official record key
device-enforcement:Z-1645-2023
Official wording
Reason: EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
Code information: UDI: (01)00884838091412 Software Version A.00.02 and A.00.01
Distribution pattern: Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
Derived failure modes
-
Unknown
EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.