Skip to content
Recall Observatory FDA recall evidence

Device product

Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.

Z-1498-2022

June 28, 2022

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 90559
Status
Ongoing
Classification
Class II
Quantity
1614 units US ; 7,979 units OUS
Official record key
device-enforcement:Z-1498-2022

Official wording

Reason: Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

Code information: Unique Device Identification (UDI): 00630414596402 All lots

Distribution pattern: Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0