Incisive CT Plus-Computed Tomography X-Ray System Model: 728149

Z-2261-2023

April 25, 2023

Class II

Product summary

Firm
Philips North America
Event
Event 92202
Status
Ongoing
Classification
Class II
Quantity
6 units OUS
Official record key
device-enforcement:Z-2261-2023

Official wording

Reason: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

Code information: UDI-DI: (01)00884838103474(21) Serial Numbers: 530408 530437 530474 530320 530332 530430

Distribution pattern: Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down