Skip to content
Recall Observatory FDA recall evidence

Device product

COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

Z-2138-2023

May 31, 2023

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 92419
Status
Ongoing
Classification
Class II
Quantity
335, 448 units
Official record key
device-enforcement:Z-2138-2023

Official wording

Reason: Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.

Code information: UDI DI 150995901HEMDILUENT5R; Container Number: 0001-0401: Lot numbers: Lot Numbers Beginning with 3541: 3541500, 3541510, 3541520, 3541550, 3541610, 3541620, 3541670, 3541690, 3541750, 3541770, 3541790, 3541810, 3541830, 3541860, 3541880, 3541900, 3541910, 3541920, 3541930, 3541940, 3541950, 3541990; Lot Numbers Beginning with 3542: 3542010, 3542030, 3542050, 3542070, 3542110, 3542150, 3542170, 3542180, 3542190, 3542230, 3542250, 3542260, 3542270, 3542310, 3542350, 3542370, 3542450, 3542490, 3542510, 3542530, 3542570, 3542590, 3542610, 3542630, 3542650, 3542670, 3542720, 3542740, 3542760, 3542780, 3542800, 3542820, 3542840, 3542850, 3542900, 3542920, 3542940, 3542960, 3542980; Lot Numbers Beginning with 3543: 3543000, 3543020, 3543040, 3543050, 3543080, 3543120, 3543140, 3543160, 3543170, 3543180, 3543200, 3543220, 3543240, 3543280, 3543300, 3543320, 3543340, 3543360, 3543370, 3543390, 3543410, 3543430, 3543460, 3543480, 3543500, 3543540, 3543560, 3543580, 3543610, 3543850, 3543870, 3543890, 3543910, 3543930, 3543950, 3543970, 3543990; Lot Numbers Beginning with 3544: 3544010, 3544030, 3544040, 3544050, 3544070, 3544090, 3544110, 3544130, 3544150, 3544160, 3544170, 3544190, 3544210, 3544230, 3544250

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.