Device product
Voalte Patient Safety, Part Number 4.0, V4.0.000 to 4.0.401, used with the Centrella Bed
Z-2006-2023
Product summary
- Event
- Event 92300
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 8 software versions
- Official record key
device-enforcement:Z-2006-2023
Official wording
Reason: Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.
Code information: UDI/DI 00887761995086, Software Versions: 4.0: 4.0.000, 4.0.100, 4.0.110, 4.0.200, 4.0.300, 4.0.301, 4.0.400, 4.0.401
Distribution pattern: US Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.
Derived failure modes
-
Unknown
Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.