Device product
FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929
Z-0367-2023
Product summary
- Firm
- Full Vision Inc
- Event
- Event 90995
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 3 units
- Official record key
device-enforcement:Z-0367-2023
Official wording
Reason: This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Code information: UDI/DI 00860176000637, Serial Numbers: FVDC-7641, FVDC-7642, FVDC-7643
Distribution pattern: Wi, GA, NY, Mexico
Derived failure modes
-
Unknown
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.