Device product
INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE
Z-2392-2023
Product summary
- Event
- Event 92664
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 12
- Official record key
device-enforcement:Z-2392-2023
Official wording
Reason: The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.
Code information: UDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003
Distribution pattern: US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA
Derived failure modes
-
Unknown
The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.