Device product
Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8.5 (G57692)
Z-1427-2023
Product summary
- Event
- Event 92083
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 3.053 units
- Official record key
device-enforcement:Z-1427-2023
Official wording
Reason: There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Code information: C-PTIS-100-HC-G-EU-FLEX7.5 (G57695): UDI-DI 00827002576957, Lot Number 14842940; C-PTIS-100-HC-G-EU-FLEX8.5 (G57696): UDI-DI 00827002576964, Lot Numbers 14727192, 14332964, 14335896, 14380758, 14380755, 14332978, 14332973, 14335892, 14335893, 14464000, 14332985, 14464008, 14463988, 14503228, 14503224, 14503222, 14503229, 14503227, 14742869, 14742870, 14727207, 14742873; C-PTIS-100-HC-G-NA-FLEX7.5 (G57691): UDI-DI 00827002576919, Lot Numbers 15063922, 15063923, 15103482, 14227740, 14221180, 14204927, 14227743, 14221125, 14221198, 14221190, 14163328, 14250265, 14250269, 14317245, 14317213, 14345634, 14376893, 14633088, 14633089, 14794549, 14794548, 14794547, 14874486, 14874484, 14926500, 14967737, 14926505, 14967739, 14967741, 14221180; C-PTIS-100-HC-G-NA-FLEX8.5 (G57692): UDI-DI 00827002576926, Lot Numbers 14608362, 14750607, 14894816, 14264399, 14299799, 14299802, 14299803, 14345667, 14365693, 14365699, 14365696, 14365694, 14365701, 14345687, 14345587, 14456438, 14456443, 14456431, 14407846, 14750610, 14488394, 14488418, 14608361, 14608363, 14750618, 14608360, 14961742, 14961716, 15089037, 15089038, 15136113
Distribution pattern: Worldwide distribution - US Nationwide.
Derived failure modes
-
Unknown
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.