Skip to content
Recall Observatory FDA recall evidence

Device product

Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

Z-2609-2023

August 03, 2023

Class II

Product summary

Firm
Randox Laboratories Ltd.
Event
Event 92912
Status
Ongoing
Classification
Class II
Quantity
66 kits
Official record key
device-enforcement:Z-2609-2023

Official wording

Reason: there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.

Code information: UDI-DI: 05055273208986; Lot Number: 1308UE

Distribution pattern: US States: GA, OH, OR, PA, TX, VA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.