Skip to content
Recall Observatory FDA recall evidence

Device product

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Z-0722-2024

December 07, 2023

Class II

Product summary

Firm
Cook Medical Llc
Event
Event 93673
Status
Ongoing
Classification
Class II
Quantity
14 US; 101 OUS
Official record key
device-enforcement:Z-0722-2024

Official wording

Reason: The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Code information: UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316

Distribution pattern: Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.