Skip to content
Recall Observatory FDA recall evidence

Device product

BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR Catalog Numbers: 546308000 (Size 8); 546310000 (Size 10); 546312000 (Size 12); 546314000 (Size 14); 546316000 (Size 16); 546318000 (Size 18); 546320000 (Size 20)

Z-0951-2023

December 19, 2022

Class II

Product summary

Firm
Depuy Ireland UC
Event
Event 91261
Status
Ongoing
Classification
Class II
Quantity
2199 US; 43515 OUS
Official record key
device-enforcement:Z-0951-2023

Official wording

Reason: All lots of BIOSTOP G Bioresorbable Cement Restrictor are being removed as a precautionary measure because tested endotoxin levels were higher than recommended by the current FDA regulatory guidance. In general, endotoxins have a potential to initiate inflammatory responses, ranging from a mild fever to potentially impact or damage to vital organs.

Code information: Catalog Number Part Description UDI/DI * 546308000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 8 10603295174547 546310000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 10 10603295174554 546312000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 12 10603295174561 546314000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 14 10603295174578 546316000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 16 10603295174585 546318000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 18 10603295174592 546320000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 20 10603295174608 * UDI-DI on GS1 0603295a0035792 Expiry dates December 2022 thru May 2025 ALL LOTS

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hungary, Ireland, Israel, Italy, Latvia, Lebanon, Luxemburg, Mexico, Netherlands, New Zealand, Norway, Portugal, Reunion, Romania, Russia, Slovakia, South Africa, Spain, Sweden, Switzerland, and Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    All lots of BIOSTOP G Bioresorbable Cement Restrictor are being removed as a precautionary measure because tested endotoxin levels were higher than recommended by the current FDA regulatory guidance. In general, endotoxins have a potential to initiate inflammatory responses, ranging from a mild fever to potentially impact or damage to vital organs.