Skip to content
Recall Observatory FDA recall evidence

Device product

IPS e.max ZirCAD CER/in. Prime BL1 C17/3- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758434

Z-1703-2024

April 05, 2024

Class II

Product summary

Firm
IVOCLAR VIVADENT AG
Event
Event 94401
Status
Ongoing
Classification
Class II
Quantity
2286 units
Official record key
device-enforcement:Z-1703-2024

Official wording

Reason: Defective blocks may show cracks after sintering If cracks are unnoticed, restorations may fracture in the patient's mouth and require restoration

Code information: UDI-DI: (01)07615208438703 (11)231117 (10)Z063MW (241)758434 (01)07615208438703 (11)231214 (10)Z067JD(241)758434 Lot Numbers: Z063MW, Z067JD

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Defective blocks may show cracks after sintering If cracks are unnoticed, restorations may fracture in the patient's mouth and require restoration