Device product

NextAR Shoulder Surgical Technique, Shoulder NextAR RSA Platform, CT based computer-assisted surgical navigation platform used to perform a reverse shoulder arthroplasty on the glenoid

Z-0372-2023

October 27, 2022

Class II

Product summary

Firm
Medacta Usa Inc
Event
Event 91091
Status
Terminated
Classification
Class II
Quantity
14 units
Official record key
device-enforcement:Z-0372-2023

Official wording

Reason: A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires adequate coracoid bone stability. Osteoporosis, weakness of the coracoid process, or other deficiencies, could lead to adverse events including bone fracture. Verify that the coracoid bone and relevant patient conditions are suitable before proceeding.

Code information: Document Reference Code 99.81NEXTAR.12US (for US market only), Revision Numbers: Rev.02 and previous

Distribution pattern: AL, FL, NH, CO, IL, SC, AR

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires adequate coracoid bone stability. Osteoporosis, weakness of the coracoid process, or other deficiencies, could lead to adverse events including bone fracture. Verify that the coracoid bone and relevant patient conditions are suitable before proceeding.