Skip to content
Recall Observatory FDA recall evidence

Device product

Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery

Z-1574-2023

November 19, 2021

Class II

Product summary

Firm
Medtronic Sofamor Danek USA Inc
Event
Event 92142
Status
Ongoing
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-1574-2023

Official wording

Reason: Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

Code information: UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103

Distribution pattern: US: IN, CT, NE

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.