Skip to content
Recall Observatory FDA recall evidence

Device product

VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Z-0728-2023

November 03, 2022

Class II

Product summary

Firm
Icotec Ag
Event
Event 91174
Status
Ongoing
Classification
Class II
Quantity
6 devices
Official record key
device-enforcement:Z-0728-2023

Official wording

Reason: During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.

Code information: UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045

Distribution pattern: US Nationwide distribution in the state of GA.

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    out of specification