Device product
Intelli-C, Left side suspended Tabletop, REF: 03400010
Z-1669-2023
Product summary
- Firm
- NRT X-RAY A/S
- Event
- Event 92230
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 13 units
- Official record key
device-enforcement:Z-1669-2023
Official wording
Reason: There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Code information: UDI-DI:05713464000121/Serial Numbers: 647009, 651002, 651006, 666009, 682003
Distribution pattern: US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Derived failure modes
-
Unknown
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.