Device product

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Z-1471-2023

April 05, 2023

Class II

Product summary

Firm
Draeger Medical Systems, Inc.
Event
Event 91938
Status
Ongoing
Classification
Class II
Quantity
6803 systems
Official record key
device-enforcement:Z-1471-2023

Official wording

Reason: The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Code information: UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to