Device product

CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384

Z-1294-2023

February 27, 2023

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 91842
Status
Ongoing
Classification
Class II
Quantity
1,168 untits
Official record key
device-enforcement:Z-1294-2023

Official wording

Reason: Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009

Code information: GTIN Number: 00889024283268 All lots

Distribution pattern: AL AZ CO FL LA MI OH OK TX Foreign: Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009