Skip to content
Recall Observatory FDA recall evidence

Device product

OEC 9800 Systems with 9-inch Image Intensifier

Z-2276-2023

May 15, 2023

Class II

Product summary

Firm
GE OEC Medical Systems, Inc
Event
Event 92504
Status
Ongoing
Classification
Class II
Quantity
58
Official record key
device-enforcement:Z-2276-2023

Official wording

Reason: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.

Code information: UDI-DI: 00840682114349. Serial Numbers: 89-0317, 89-0554, 8S-1293, 8S-0509, 8S-0108, 8S-1029, 8S-2836, 89-3028, 8S-1574, 89-0609-RC, 89-1308, 8S-0394, 89-0979, 8SXXXX03626, 8SXXXX03683, 8SXXXX03681, 8S-1894, 8S-0175, 8S-2988, 89-2459, 89-2101, 89-3245, 89-1623, 89-0296, 8S-0200, 8S-0748, 89-1590, 89-1713, 8S-0124, 89-0282, 8S-2580, 89-3132, 8S-3039, 89-3451, 89-0223, 89-1155, 8SXXXX03680, 8S-3278, 8S-2426, 89-2913, 8S-1167, 89-1278, 89-1854, 8S-0233-C, 8S-1926, 8SXXXX03682, 89-2167, 8S-1109, 8SXXXX03679, 8S-0410, 89-1579-N, 89-1920, 89-3102, 8S-2046-C, 89-2340, 89-0461, 89-3809, 89-3810

Distribution pattern: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.