Device product
Medex LOGICAL CATH LAB KIT, List Number M20754
Z-0739-2024
Product summary
- Event
- Event 93506
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 31,685 total
- Official record key
device-enforcement:Z-0739-2024
Official wording
Reason: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Code information: UDI-DI: 10351688510686; Lot Number: 4371054
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, UAE.
Derived failure modes
-
Labeling or packaging
inadequate seal
-
Manufacturing or process control
manufacturing defect