Device product
Puritan Bennett 980 Series Ventilator, 980A1ENNISB
Z-0338-2022
Product summary
- Firm
- Covidien Llc
- Event
- Event 88987
- Status
- Terminated
- Classification
- Class I
- Quantity
- 278 total
- Official record key
device-enforcement:Z-0338-2022
Official wording
Reason: Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Code information: GTIN/UDI - 10884521201897 Serial Numbers: 35B1700507, 35B1700533, 35B1700539, 35B1700465, 35B1700580, 35B1700530, 35B17005363, 5B1700542, 35B1700509, 35B1701614, 35B1700527, 35B1700537, 35B1700541, 35B1700508 & 35B1701569.
Distribution pattern: Worldwide distribution - US Nationwide.
Derived failure modes
-
Manufacturing or process control
Manufacturing assembly error