Device product
CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US
Z-0939-2022
Product summary
- Event
- Event 89810
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-0939-2022
Official wording
Reason: Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Code information: All lots UDI: 03700506307319
Distribution pattern: US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Derived failure modes
-
Potency or specification failure
does not meet the specification