Skip to content
Recall Observatory FDA recall evidence

Device product

Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system

Z-0776-2023

November 21, 2022

Class II

Product summary

Firm
Medtronic MiniMed
Event
Event 91260
Status
Ongoing
Classification
Class II
Quantity
3235
Official record key
device-enforcement:Z-0776-2023

Official wording

Reason: An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not receive SMS notifications (could result in hypoglycemia or hyperglycemia), and sensor glucose values will also not be sent to the InPen app.

Code information: App Version 1.3.0

Distribution pattern: OUS: Austria, Belgium, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Luxembourg, Moldova, Republic of, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not receive SMS notifications (could result in hypoglycemia or hyperglycemia), and sensor glucose values will also not be sent to the InPen app.