Device product
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM
Z-0423-2023
Product summary
- Event
- Event 91127
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 129,086
- Official record key
device-enforcement:Z-0423-2023
Official wording
Reason: The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Code information: ALL serial numbers/ controller lots are currently in scope of this recall
Distribution pattern: Domestic distribution nationwide.
Derived failure modes
-
Unknown
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.