Skip to content
Recall Observatory FDA recall evidence

Device product

Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable

Z-2664-2023

August 10, 2023

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 92893
Status
Ongoing
Classification
Class II
Quantity
60102 units
Official record key
device-enforcement:Z-2664-2023

Official wording

Reason: Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Code information: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX 00815686020613; INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx 00815686020590; INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx 00815686020576; INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx 00815686020538; INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" 00815686020538; INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC 00815686020477; INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm 00815686020446; INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm 00815686020439; INS-0050 vPad Patient Tracker, 5/Bx 00815686020408; INS-0048 vPad Connector Cable 00815686020392; All serial numbers and all lot numbers

Distribution pattern: Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment