Device product
Puritan Bennett 980 Series Ventilator, 980X1ENNISB
Z-0348-2022
Product summary
- Firm
- Covidien Llc
- Event
- Event 88987
- Status
- Terminated
- Classification
- Class I
- Quantity
- 278 total
- Official record key
device-enforcement:Z-0348-2022
Official wording
Reason: Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Code information: GTIN/UDI - 10884521171749 Serial Numbers: 35B1700430 35B1700442 35B1700456 35B1700510 35B1700514 35B1700520 35B1700553 35B1700554 35B1700559 35B1700561 35B1700608
Distribution pattern: Worldwide distribution - US Nationwide.
Derived failure modes
-
Manufacturing or process control
Manufacturing assembly error