Device product
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Z-0040-2024
Product summary
- Firm
- AB Sciex, LLC
- Event
- Event 92810
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 14
- Official record key
device-enforcement:Z-0040-2024
Official wording
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Code information: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
Distribution pattern: US Nationwide.
Derived failure modes
-
Unknown
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.