Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Z-0040-2024

August 30, 2023

Class II

Product summary

Firm
AB Sciex, LLC
Event
Event 92810
Status
Ongoing
Classification
Class II
Quantity
14
Official record key
device-enforcement:Z-0040-2024

Official wording

Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Code information: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203

Distribution pattern: US Nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.