Skip to content
Recall Observatory FDA recall evidence

Device product

Access Thyroglobulin Calibrators

Z-0100-2024

July 06, 2023

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 92670
Status
Ongoing
Classification
Class II
Quantity
5980 units
Official record key
device-enforcement:Z-0100-2024

Official wording

Reason: It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Code information: UDI/DI 15099590227166, Serial/Lot Numbers: 233806, 234171, 234355, 234443, 234444, 234617, 337940

Distribution pattern: worldwide except US, Canada, and China

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.