Skip to content
Recall Observatory FDA recall evidence

Device product

ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)

Z-2666-2023

August 08, 2023

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 92971
Status
Ongoing
Classification
Class II
Quantity
4966 reagents
Official record key
device-enforcement:Z-2666-2023

Official wording

Reason: There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias will vary wedge to wedge.

Code information: (6 x 350 tests) UDI-DI: 00630414561974; (7 x 145 tests) UDI-DI: 00630414513157; Lot Numbers: 612584 625096 612579 625091 635826 635821

Distribution pattern: US Nationwide. Global Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias will vary wedge to wedge.