Device product
FreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 Operating System
Z-0734-2024
Product summary
- Event
- Event 93393
- Status
- Completed
- Classification
- Class II
- Quantity
- 45,173
- Official record key
device-enforcement:Z-0734-2024
Official wording
Reason: If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required.
Code information: UDI-DI: 00357599816001, Software v.3.4.0 and below
Distribution pattern: US Nationwide distribution in the states of NY, CA, NJ, UT, PA, MS, OH, TX, MN, FL, LA, NE, MI, VA, NC, MA, IN, SC, IA.
Derived failure modes
-
Unknown
If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required.