Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC

Z-0717-2024

December 04, 2023

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 93613
Status
Ongoing
Classification
Class II
Quantity
236 units
Official record key
device-enforcement:Z-0717-2024

Official wording

Reason: The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.

Code information: Lot Number: All units sold prior to 2023

Distribution pattern: US Nationwide. Global Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.