DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

Z-0107-2024

September 12, 2023

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 93023
Status
Ongoing
Classification
Class II
Quantity
36 units
Official record key
device-enforcement:Z-0107-2024

Official wording

Reason: The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.

Code information: UDI/DI 15099590732103, All DxI 9000 Access Immunoassay Analyzer running software version 1.15 or prior

Distribution pattern: Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.