Skip to content
Recall Observatory FDA recall evidence

Device product

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01

Z-0093-2024

September 06, 2023

Class II

Product summary

Firm
CareFusion 303, Inc.
Event
Event 93016
Status
Ongoing
Classification
Class II
Quantity
61
Official record key
device-enforcement:Z-0093-2024

Official wording

Reason: When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.

Code information: REF: 1116-00, UDI-DI: 10885403512605. Software Versions: 1.7.3 and 1.7.4 (Limited commercial release)

Distribution pattern: US Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.