Skip to content
Recall Observatory FDA recall evidence

Device product

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

Z-1181-2024

February 08, 2024

Class II

Product summary

Firm
PTW-FREIBURG
Event
Event 93988
Status
Ongoing
Classification
Class II
Quantity
1 US; 65 worldwide
Official record key
device-enforcement:Z-1181-2024

Official wording

Reason: Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.

Code information: UDI-DI: EPTWS070031 Software VERIQA (S070031), version 2.0 and 2.1

Distribution pattern: GA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.