Device product
Revolution CT
Z-2004-2021
Product summary
- Event
- Event 88150
- Status
- Completed
- Classification
- Class II
- Quantity
- 197 (US = 40; OUS = 157)
- Official record key
device-enforcement:Z-2004-2021
Official wording
Reason: GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.
Code information: 5590000 5590000-6 5590000-2
Distribution pattern: Worldwide distribution - US Nationwide
Derived failure modes
-
Unknown
GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.