ARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35

Z-1479-2024

March 04, 2024

Class II

Product summary

Firm
Abbott GmbH
Event
Event 94205
Status
Ongoing
Classification
Class II
Quantity
7219 units
Official record key
device-enforcement:Z-1479-2024

Official wording

Reason: Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).

Code information: a) 1L79-25, 00380740001414, Lot Numbers: 52011BE00, 54140BE00, 56375BE00, 58163BE00, 59194BE00; b) 1L79-35, 00380740001421, Lot Numbers: 52015BE00, 54144BE00, 56379BE00, 58167BE00, 59198BE00

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).