8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System
Z-0123-2024
Product summary
- Event
- Event 92978
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 9771
- Official record key
device-enforcement:Z-0123-2024
Official wording
Reason: Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
Code information: UDI-DI: 04046964660887 Serial Numbers: 868497 -892669 ***Updated 8/7/24*** SNs added: 430056, 430119, 430155, 430243, 551518, 685947, 703196, 705968, 726936, 737729, 753364, 774163, 790843, 814074
Distribution pattern: Domestic distribution nationwide. Foreign distribution to Canada.
Derived failure modes
-
Unknown
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.