Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Z-1594-2024

January 31, 2024

Class II

Product summary

Firm
Randox Laboratories Ltd.
Event
Event 94138
Status
Ongoing
Classification
Class II
Quantity
6 units
Official record key
device-enforcement:Z-1594-2024

Official wording

Reason: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Code information: GTIN: 05055273204544 (1) Batch/Lot: 602114 /1083MA-1087MA (2) Batch/Lot: 636645 /1108MA-1112MA (3) Batch/Lot: 662429 /1083MA- 1087MA

Distribution pattern: US Nationwide distribution in the states of GA, MA, WV including PR.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.