Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

Z-1172-2024

January 03, 2024

Class II

Product summary

Firm
FUJIFILM Healthcare Americas Corporation
Event
Event 93855
Status
Ongoing
Classification
Class II
Quantity
8
Official record key
device-enforcement:Z-1172-2024

Official wording

Reason: The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Code information: UDI-DI 00854904006015 Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.0

Distribution pattern: US domestic distribution nationwide. International distribution to Israel.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.