Device product
Monaco Radiation Treatment Planning (RTP) System
Z-0965-2019
Product summary
- Firm
- Elekta, Inc.
- Event
- Event 82072
- Status
- Terminated
- Classification
- Class II
- Quantity
- 14 devices
- Official record key
device-enforcement:Z-0965-2019
Official wording
Reason: If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco¿ will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.
Code information: Software version 5.40
Distribution pattern: Domestic distribution to Wisconsin, Texas. International distribution to Denmark, England, Germany, Netherlands, Sweden.
Derived failure modes
-
Unknown
If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco¿ will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.