Skip to content
Recall Observatory FDA recall evidence

Device product

StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.

Z-2707-2020

June 05, 2020

Class II

Product summary

Firm
Medtronic Navigation, Inc.
Event
Event 85901
Status
Terminated
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-2707-2020

Official wording

Reason: There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

Code information: StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.

Distribution pattern: US Nationwide distribution including in the states of Alabama and Massachusetts.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.