Device product

AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332

Z-2581-2021

August 16, 2021

Class II

Product summary

Firm
Biodex Medical Systems, Inc.
Event
Event 88604
Status
Terminated
Classification
Class II
Quantity
429 units
Official record key
device-enforcement:Z-2581-2021

Official wording

Reason: When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration

Code information: Serial Number Range:17080651-21061047 UDI: 00718175003329

Distribution pattern: Nationwide Foreign: Austria Autralia Brazil CANADA Chile Colombia Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt France Hong Kong HUNGARY Iceland India Iraq Ireland ISRAEL Isreal Italy Japan JORDAN Kuwait Lebanon Libya Malaysia MALTA Mexico Morocco Netherlands New Zealand Oman Pakistan Palestine Panama Peru Philippines Poland Portugal PR Puerto Rico Qatar Romania Russia Saudi Arabia Singapore South Africa South Korea Spain Sweden Taiwan UAE United Arab Emirates United Kingdom Venezuela

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may