Device product
Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe
Z-1157-2024
Product summary
- Event
- Event 93795
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 4831 units
- Official record key
device-enforcement:Z-1157-2024
Official wording
Reason: Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).
Code information: UDI/DI 40080196031765 (case), 10080196031764 (unit) Lot Numbers: 96922110001 and 96923100001
Distribution pattern: US, Nationwide
Derived failure modes
-
Sterility assurance
sterile barrier to be breached