Skip to content
Recall Observatory FDA recall evidence

Device product

Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx Single Dip Card Single Drug Detection: Buprenorphine (BUP)", X11-IS1 BAR-DX /"Alere iScreen" Dx Single Dip Card Single Drug Detection: Barbiturate (BAR)", X11-IS1 OXY-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Oxycodone (OXY)", X11-IS1 BZO-DX / "Alere iScreen" Dx Single Dip Card Single Drug Detection: Benzodiazepine (BZO300)", X11-IS1 MET-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methamphetamine (MET)", X11-IS1 AMP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Amphetamine (AMP)", X11-IS1 MDM-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Ecstasy (MDMA)", X11-IS1 MTD-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methadone (MTD)", X11-IS1 MOP-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Morphine (MOP)", X11-IS1 THC-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Marijuana (THC)", X11-IS1 TCA-DX/ "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Tricyclic Antidrepessants (TCA)", X11-IS1 PCP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Phencyclidine (PCP)",

Z-3029-2020

August 18, 2020

Class II

Product summary

Firm
Ameditech Inc
Event
Event 86335
Status
Terminated
Classification
Class II
Quantity
303,100 units
Official record key
device-enforcement:Z-3029-2020

Official wording

Reason: Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Code information: ALL Lots Within Expiry

Distribution pattern: US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.