Device product

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Z-0658-2022

December 17, 2021

Class II

Product summary

Firm
Nobel Biocare Usa Llc
Event
Event 89465
Status
Terminated
Classification
Class II
Quantity
702 pieces
Official record key
device-enforcement:Z-0658-2022

Official wording

Reason: Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Code information: Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Distribution pattern: Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.