Device product
Beckman Coulter Lipase, REF OSR6130 and OSR6230.
Z-0210-2020
Product summary
- Event
- Event 83583
- Status
- Terminated
- Classification
- Class II
- Quantity
- 20,926 units
- Official record key
device-enforcement:Z-0210-2020
Official wording
Reason: NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.
Code information: All lots
Distribution pattern: Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.
Derived failure modes
-
Unknown
NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.