Skip to content
Recall Observatory FDA recall evidence

Device product

Access Substrate, REF 81906, For use with the Access Immunoassay Systems

Z-1376-2024

February 07, 2024

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 94123
Status
Ongoing
Classification
Class II
Quantity
725,568 units (Updated)
Official record key
device-enforcement:Z-1376-2024

Official wording

Reason: On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman Coulter has identified that Access Substrate bottles in the identified lots were released from manufacturing without labels. Missing label may cause some patient results reporting delay but will not cause any errorreous results. However, because labs are required to have redundancy such that delay would not significantly impact clinical management and because the defect is easily recognized by the user, the probability of harm is remote.

Code information: UDI/DI 15099590229986, Lot Numbers: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, 338538. UPDATED - ADDITIONAL LOT NUMBERS PER EXPANSION EMAIL DATED 10/30/2024: 234600, 234633, 234682, 234706, 234721, 234759, 234790, 234801, 234834, 234862, 337904, 337930, 337935, 337959, 337984, 338040, 338065, 338119, 338163, 338214, 338229, 338254, 338270, 338289, 338322, 338333, 338346, 338420, 338435, 33845, 33848, 338511, 338538, 338566, 338587, 338621, 338622, 338698, 338719, 338720, 338738, 338753, 338762, 338783, 338860, 338996, 339040, 339041, 339056, 339107, 339106, 339146, 439113, 439111, 439157, 439229, 439239, 439285¿.

Distribution pattern: Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Lithuania, Macao, Malaysia, Maldives, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Oman, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen, Zambia UPDATE PER EXPANSION EMAIL DATED 10/30/2024: Distribution was made nationwide, including Guam and Puerto Rico. There was government distribution but no military distribution. The following countries received affected product identified for the recall expansion: Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Costa Rica, C¿te d'Ivoire, Croatia, Czechia, Denmark, Ecuador, Egypt, El Salvador, Estonia, Eswatini, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Macao, Malawi, Malaysia, Maldives, Mauritius, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uruguay, Uzbekistan, Viet Nam, Yemen, Zambia, Zimbabwe.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    Missing label