Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part

Z-1543-2021

February 26, 2021

Class I

Product summary

Firm
STEP-HAR MEDICAL LLC
Event
Event 87773
Status
Terminated
Classification
Class I
Quantity
2
Official record key
device-enforcement:Z-1543-2021

Official wording

Reason: Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Code information: Serial Numbers: 12747983 and 9919591

Distribution pattern: IA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.